Model Number 20226
Medical Device Problem Codes Difficult to Remove (1528); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: used (b)(6) 2020 as the information available is (b)(6) 2020 only.
 
Event or Problem Description
It was reported that removal difficulties and stent dislodgement occurred, requiring additional intervention.The target lesion was located at the right iliac vein.A 8.0x40x75cm express ld stent was advanced for treatment.A picture was taken but could not see the vessel well, so the stent was pulled back; however, resistance was felt.During fluoroscopy, it was noticed that the stent came off of the balloon.The sheath, balloon and stent were pulled back in one unit.The sheath and balloon were out of the body and it was noted that the stent was left in the femoral artery right at the access site.Additional lidocaine was given, and a small nick was made in the skin to pull the stent out using the hemostat.The procedure was continued and a new sheath was inserted in the left femoral artery.A new stent was advanced and deployed without complications.The patient was stable during the whole procedure and was sent to recovery after.The patient was discharged on the same day.
 
Event or Problem Description
It was reported that removal difficulties and stent dislodgement occurred, requiring additional intervention.The target lesion was located at the right iliac vein.A 8.0x40x75cm express ld stent was advanced for treatment.A picture was taken but could not see the vessel well, so the stent was pulled back; however, resistance was felt.During fluoroscopy, it was noticed that the stent came off of the balloon.The sheath, balloon and stent were pulled back in one unit.The sheath and balloon were out of the body and it was noted that the stent was left in the femoral artery right at the access site.Additional lidocaine was given, and a small nick was made in the skin to pull the stent out using the hemostat.The procedure was continued and a new sheath was inserted in the left femoral artery.A new stent was advanced and deployed without complications.The patient was stable during the whole procedure and was sent to recovery after.The patient was discharged on the same day.
 
Additional Manufacturer Narrative
B3 - date of event: initial reporter confirmed event date was (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10442933
Report Number2134265-2020-11557
Device Sequence Number2713889
Product Code NIO
UDI-Device Identifier08714729392279
UDI-Public08714729392279
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2022
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0023765606
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/18/2020
Supplement Date Received by Manufacturer09/01/2020
Initial Report FDA Received Date08/24/2020
Supplement Report FDA Received Date09/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
-
-