| Model Number |
ECR45M |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #unk.As the device was not returned, an analysis investigation could not be performed.A conclusion could not be reached as to what may have caused or contributed to the event.Complaint information is trended on a regular basis to determine if further investigation is warranted.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, after firing, some staples formed with irregular shapes and the anastomotic site was oozing.Another two reloads had the same issue.Oozing was stopped with compression hemostasis.There was no patient consequence reported.No additional information is available at this time.
|
| |
|