• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AUGMEDICS LTD XVISION SPINE SYSTEM (XVS); NAVIGATION SYSTEM WITH AUGMENTED REALITY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AUGMEDICS LTD XVISION SPINE SYSTEM (XVS); NAVIGATION SYSTEM WITH AUGMENTED REALITY Back to Search Results
Model Number AMCH00005
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Spinal Column Injury (2081)
Date of Event 07/28/2020
Type of Reportable Event Serious Injury
Event or Problem Description
The event is related to the procedure performed on (b)(6) 2020.The surgeon reported that 3 of the screws deviated in position.The first screw inserted was accurate.During the investigation it was found out the surgeon was working though small incision in open procedure which caused for constant rearrangements of retractors and instrumentation handling in close vicinity of the clamp.The augmedics representative suspects that the clamp closing screw was not fully tightnened.It was concluded that the clamp was not tighten enough by the surgeon and therefore was moved by the handling of the retractor or instrumentation the 3 screws were removed and replaced using fluoro.No additional intervention was required.No additional adverse events were reported.The surgeon was re-trained on indication for fully closed clamp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XVISION SPINE SYSTEM (XVS)
Common Device Name
NAVIGATION SYSTEM WITH AUGMENTED REALITY
Manufacturer (Section D)
AUGMEDICS LTD
1 ha-tsmikha st
yokneam illit, 20692 05
IS  2069205
MDR Report Key10446961
Report Number3016571711-2020-00001
Device Sequence Number14464006
Product Code OLO
UDI-Device Identifier07290113780002
UDI-Public(01)07290113780002(11)200402
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190929
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAMCH00005
Device Catalogue NumberAMCH00005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2020
Initial Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient Weight61
-
-