• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL GUIDE HANDLE ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL GUIDE HANDLE ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674001
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 07/30/2020
Event Type  malfunction  
Event Description
It was reported that the sealing material on the drill guide handle is cracking. No case involved.
 
Manufacturer Narrative
Results of investigation: the device, intended for use in treatment, was not returned for evaluation. Therefore, product analysis could not be performed at this time. So the reported event could not be confirmed. A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions. This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur. Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event. We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary. At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture. Based on this investigation, the need for corrective action is not indicated. Without the actual product involved and/or device information, our investigation cannot proceed. If the device or new information is received in the future, this complaint can be re-opened. No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary. We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDRILL GUIDE HANDLE
Type of DeviceROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
sarah freestone
1450 brooks rd.
memphis, TN 38116
0447940038
MDR Report Key10447657
Report Number1020279-2020-04178
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560469
UDI-Public03596010560469
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number71674001
Device Catalogue Number71674001
Device Lot Number4895315403321
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/09/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
-
-