• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 670185
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/07/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the cuff component of the bivona tracheostomy was found to be leaking after 72 hours of use.No patient injury or complications were reported in relation to this event.
 
Additional Manufacturer Narrative
Other, other text: device evaluation- two used devices were returned for evaluation.The returned devices were given leak testing; this showed leakage at the same area for both devices.The leak site was determined to be a cut area on the cuffs located approximately 25mm from the distal tip.Because the device is 100% leak tested prior to packaging it is unlikely this damage was present during manufacturing.The cuff damage was present on both samples in same area and were returned for separate events for the same patient (related file cc-0083933.It was determined likely the issue occurred when the cartilage in the patient's trachea rubbed against the cuff surface and eventually caused the cut damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
MDR Report Key10449253
Report Number3012307300-2020-08442
Device Sequence Number4031040
Product Code JOH
UDI-Device Identifier15021312006087
UDI-Public15021312006087
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number670185
Device Catalogue Number670185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/27/2020
Supplement Date Received by Manufacturer11/25/2020
Initial Report FDA Received Date08/25/2020
Supplement Report FDA Received Date12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-