| Model Number |
0375708500 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported there was a thermal event.Please note this was noticed during inspection.
|
| |
|
Event or Problem Description
|
|
It was reported there was a thermal event.Please note this was noticed during inspection.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received at the local service facility for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.The technical service report is attached (see comm log), and indicates: "fault: faulty keypad and motor over amp during inspection of the shaver.Repair level: level 1.Action taken: replacement of defective motor and keypad revision of the shaver.Tests: function test general inspection qip." probable root cause: 1.Material/design error.2.Manufacturing/assembly error (missing lubrication/grease).3.Clogged suction path.4.Hot irrigant fluid external to shaver.5.Use error.
|
| |
|