• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. AHTB ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. AHTB ACTIVE HEEL TRACTION BOOT; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 72202682
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the active heel traction boot was not holding tight.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: h2, h3, h6: the device, which was used in a procedure, was not returned for evaluation.A relationship between the device and the reported incident could not be confirmed.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.Factors that are known to contribute to the alleged fault/failure may be shipping damage, a component failure, or wear from use.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product is returned in the future the complaint can be reopened and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB ACTIVE HEEL TRACTION BOOT
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10451663
Report Number3003604053-2020-00088
Device Sequence Number16973323
Product Code HST
UDI-Device Identifier00885554024807
UDI-Public00885554024807
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202682
Device Catalogue Number72202682
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer09/29/2020
Initial Report FDA Received Date08/25/2020
Supplement Report FDA Received Date09/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-