| Model Number |
3100 A |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire file identification: (b)(4).The reported issue was also by rt dept while doing the patient circuit calibration.As a resolution, the customer will replaced the panel meter.The suspect/device component has not been returned to vyaire.Therefore, a definitive root cause cannot be determined.Any additional information received from the customer will be included in a follow-up report.
|
| |
|
Event or Problem Description
|
|
It is reported to vyaire medical that 3100a ventilator experienced a fluctuating map between 53-58 cm on the display while checking the zero/span on this unit.The external meter is reading stable at 37 cm.The customer confirmed that there was no patient involvement associated with the reported event.
|
| |
|