G4:06apr2021.B4:07apr2021.A philips field service engineer (fse) was dispatched to the customer site and it was determined that the data acquisition printed circuit board assembly (da pcba) and flow sensor assembly needed to be replaced to return the device to working condition.The fse replaced the da pcba and flow sensor to resolve the reported issue.The device passed performance verification testing.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|