|
The returned chest drain was evaluated and there was no visible damage of the chest drain.The blue water for the water seal chamber was not present so, per the instructions for use (ifu), the water seal was re-filled to the 2cm mark with water.Set up: step 1.Fill water seal to 2 cm line ¿ hold funnel down and fill to top.Raise funnel to fill water seal to 2 cm fill line.For models available with sterile fluid, twist top off bottle and insert tip into suction port.Squeeze contents into water seal until fluid reaches 2 cm fill line.The suction control chamber¿s water column was filled to -5cmh20 level and a suction source vacuum of -109mmhg was applied.Step 2.Fill suction control chamber to desired pressure level ¿ remove vent plug; pour water to desired suction level.Replace vent plug.Step 3.Connect patient tube to patient ¿ connect chest drain to patient prior to initiating suction.Step 4.Connect suction to chest drain ¿ attach suction line to suction port on top of chest drain.Turn suction source on until constant, gentle bubbling occurs in chamber a".The stopcock was cracked open to allow slow bubbling in the suction control chamber as indicated in the ifu.Suction control stopcock: for models equipped with suction control stopcock, it can be adjusted to attain gentle bubbling in suction control chamber.Stopcock should be left in the open or on position at all times.Bubbling in the suction control chamber indicates sufficient wall source suction and the application of the suction level per the water column height.The bubbling should not be vigorous bubbling but, rather, gentle bubbling.Gentle bubbling keeps noise level down and lowers the rate of evaporation (compared to vigorous bubbling.) the float ball as expected under normal conditions was being pulled down to the +2 mark within the water seal as suction is attempting to remove any positive pressure from the system.The complaint mentioned the patient having a coughing fit.During a coughing fit, the movement of the lungs will push out any residual air between the parietal and pulmonary pleura.This creates negative pressure in the pleural space (i.E.Normal state).If negative pressure were present, the float ball would rise in the column.Since active suction was applied to the system and the patient was coughing, the pressure inside the drain (after the coughing fit) was negative (i.E.Vacuum) and this is the normal to see in the drain.Changes in patient pressure: with suction on, patient pressure will equal suction control chamber a water level plus the height of water seal column level b.For gravity drainage (no suction), patient pressure will equal the height of the graduated water seal column level only.If no active suction is operating on the drain (i.E.Gravity drainage), the movement of the water seal float ball is completely dependent upon the dynamic nature of the pleural space.In the complaint, the water seal float ball had risen indicating the presence of negative pressure (normal physiological condition of the pleural space) and the float ball stopped moving.The cessation of the float ball movement would indicate a blockage proximal to the patient (i.E.Upstream).The details also indicate that this device was used with a merit medical resolve 8.5fr pigtail drainage catheter.If this small-bore catheter becomes occluded this could also explain why the float ball was not reactive.Based on the details of the complaint and the properly functioning drain that was returned it appears as if there may be a misunderstanding of the operation of the drain.
|