• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 38275
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/10/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the catheter broke.The totally occluded target lesion was located in the calcified left common iliac artery.Access to the lesion was from the right groin.A 4fr imager ii angiograhphic catheter was advanced over a v-18 guidewire.During the procedure, the imager embedded in the chronic total occlusion (cto).The physician tried backing out the device but experienced difficulty due to the cto.The v-18 wire was switched to a different wire to see if this would help.This was not successful.In attempts to pull the imager out from the cto, the imager catheter broke approximately 80mm from the distal end.The broken section was retrieved via snare and no fragments were left in the patient.There were no further patient complications and the patient was fine.It was possible that the device was kinked and broke at this location.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: analysis showed that the device was separated approximately 11.5cm from the tip.There were multiple small bends and kinks located throughout the shaft.The catheter looked to be stretched, which would give the indication of tensile force.This damage is consistent with patient conditions such as a severely calcified, stenosed lesion and tortuous anatomy.The reported information stated that the device was imbedded in a total chronic occlusion.Tensile testing was performed on the returned unit by pulling approximately 84cm from the strain relief.Maximum tensile force was 7lbs.The minimum specification is 2.2lbs.This confirms the device met specification.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that the catheter broke.The totally occluded target lesion was located in the calcified left common iliac artery.Access to the lesion was from the right groin.A 4fr imager ii angiographic catheter was advanced over a v-18 guidewire.During the procedure, the imager embedded in the chronic total occlusion (cto).The physician tried backing out the device but experienced difficulty due to the cto.The v-18 wire was switched to a different wire to see if this would help.This was not successful.In attempts to pull the imager out from the cto, the imager catheter broke approximately 80mm from the distal end.The broken section was retrieved via snare and no fragments were left in the patient.There were no further patient complications and the patient was fine.It was possible that the device was kinked and broke at this location.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMAGER II ANGIOGRAPHIC CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key10452311
Report Number2134265-2020-11600
Device Sequence Number4044715
Product Code DQO
UDI-Device Identifier08714729404408
UDI-Public08714729404408
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K120893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2021
Device Model Number38275
Device Catalogue Number38275
Device Lot Number0000145905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2020
Initial Date Received by Manufacturer 08/12/2020
Supplement Date Received by Manufacturer09/25/2020
Initial Report FDA Received Date08/25/2020
Supplement Report FDA Received Date10/22/2020
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-