ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 1012536-100
Medical Device Problem Codes Difficult or Delayed Activation (2577); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual and functional analysis was performed on the returned device.The reported difficulties were unable to be confirmed as the stent was already fully deployed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other complaints.The investigation was unable to determine cause for the reported difficulties.It may be possible that the distal shaft was bent or restricted in the anatomy preventing the shaft lumens from moving freely and causing the resistance with the thumbwheel; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed via antegrade approach in the left common femoral artery, to treat a lesion in the mid to distal superficial femoral artery (sfa).Laser atherectomy was performed.The absolute pro vascular stent delivery system was advanced and deployment was initiated.It was noted that the thumbwheel appeared stuck.The system was locked and unlocked twice, then the handle was jiggled.The thumbwheel was able to turn; however, resistance was noted.The physician was able to turn the thumbwheel with difficulty and the stent was able to be fully deployed at the target lesion.There was no adverse patient effect or clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10452335
Report Number2024168-2020-07109
Device Sequence Number11592209
Product Code NIP
UDI-Device Identifier08717648176173
UDI-Public08717648176173
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012536-100
Device Catalogue Number1012536-100
Device Lot Number9112261
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2020
Initial Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight100
-
-