| Model Number |
MMT-751NAL |
| Medical Device Problem Code |
Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The insulin pump received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test due to blank display.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that there was fluid in insulin pump and under the screen.Customer was running and it was exposed to sweat.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|