Model Number DACAPO POWERPACK |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been explanted and has been returned to innsbruck where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
A leaking/ broken dacapo battery was received at service and repair.On the repair request form there were battery issues, the battery would not charge.
|
|
Manufacturer Narrative
|
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed, most likely due to external mechanical stress.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.This is a final report.
|
|
Event Description
|
A leaking/ broken dacapo battery was received at service and repair.On the repair request form there were battery issues, the battery would not charge.The user did not come into contact with the leaking electrolyte.
|
|