Olympus medical systems corp.
(omsc) was informed that between (b)(6) 2019 and (b)(6) 2020, after the flexible cystoscopy twelve patients were infected with pseudomonas aeruginosa likely to have been acquired from the subject device.
All of the pseudomonas aeruginosa isolates from clinical samples matched strains of pseudomonas aeruginosa isolated from the subject device.
The four samples were collected from blood, the one was collected from knee joint fluid, and the seven were collected from urine.
Also, microbe was not detected from any samples of the cystoscopes.
The six of the affected patients required admission for iv antibiotics, the one of these six patients required intensive treatment unit (itu) admission, and another required multiple surgical procedures for the septic arthritis.
Also two of the six patients died after isolating pseudomonas aeruginosa, though it was not clear if their deaths were a direct consequence of the infection as they had other co-morbidities.
All elective cystoscopy procedures have been suspended until the devices have been replaced and a suitable decontamination/reprocessing procedure is in place.
17 pieces of maj-891 (forceps/irrigation plug) were sampled in total.
16 have had bacterial growth.
Pseudomonas aeruginosa was isolated from 6 of the devices.
5 of these matched pseudomonas aeruginosa isolated from clinical samples.
The facility suspected the source might be continued pseudomonas bacteria inside the rubber bungs of maj-891 that the facility attached to the olympus' cystoscope to control water flow.
The facility stated that maj-891 did not appear to be reprocessed with the reprocessor.
Also the facility was concerned about the reprocessing method of the subject device which was provided by olympus.
According to the number of patients and microbiological testing of the subject device, omsc is submitting 28 medical device reports.
12 reports are regarding the patient¿s adverse events (death and infection).
16 reports are regarding the results for positive of the microbiological testing of the subject device.
This is 6 of 28 reports.
|