• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Olympus field service engineer (fse) confirmed that a cracked push plate was found while performing preventive maintenance.The device was repaired on site and verified that it was working as specified.The equipment passed the electrical and safety tests.If additional information is obtained a supplemental mdr will be filed.
 
Event Description
A technician reported that an endoscope reprocessor required repairs on a cracked lid found during preventive maintenance.No patient injury was reported.No additional information has been obtained.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿when closing the lid, be careful not to get the connecting tube caught between the reprocessing basin and lid and make sure the endoscopes and the washing case are not touching the lid.Close the lid after ensuring that the cover of the washing case is closed.¿ the root cause was not determined.Probable causes are that the lid was contacted by a scope when it was closed and/or something hard hit the lid causing the damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10454216
MDR Text Key204398782
Report Number8010047-2020-05898
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-