• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0500J
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the intra-aortic balloon pump (iabp) switched to battery power despite being plugged in while in use.As a result, the iabp was swapped out for another iabp.There was no report of patient complications, serious injury or death.
 
Additional Manufacturer Narrative
(b)(4).No iabp part was returned to teleflex chelmsford for investigation.The reported complaint of "the iabp was switched to be battery-powered despite it being plugged" is not able to be confirmed.The iabp was tested by a distributor's technician and the issue could not be replicated.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the iap serial and lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
 
Event or Problem Description
It was reported that the intra-aortic balloon pump (iabp) switched to battery power despite being plugged in while in use.As a result, the iabp was swapped out for another iabp.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE JAPANESE
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10454306
Report Number3010532612-2020-00240
Device Sequence Number16947935
Product Code DSP
UDI-Device Identifier00801902043429
UDI-Public00801902043429
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberIAP-0500J
Device Lot NumberN/A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/13/2020
Supplement Date Received by Manufacturer10/02/2020
Initial Report FDA Received Date08/26/2020
Supplement Report FDA Received Date10/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-