Model Number M00558390
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the physician noticed that the balloon was leaking.Reportedly, the balloon was removed from the patient and noted that the balloon had a visible pinhole.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the physician noticed that the balloon was leaking.Reportedly, the balloon was removed from the patient and noted that the balloon had a visible pinhole.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block e1: initial reporter city: (b)(6).Block h6: problem code 1504 captures the reportable event of balloon pinhole.Block h10: investigation results: visual examination of the returned complaint device found the balloon did not show visual defects and was in a good condition.No damage was found on the catheter of the device.Functional analysis was performed, and the balloon was inflated without a problem; however, the balloon would not hold pressure due to a pinhole in the body of the balloon.This failure is likely due to factors encountered during the procedure, such as handling or manipulation during the preparation, initial set-up, or procedure that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10454521
Report Number3005099803-2020-03453
Device Sequence Number3524209
Product Code FDF
UDI-Device Identifier08714729339328
UDI-Public08714729339328
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2022
Device Model NumberM00558390
Device Catalogue Number5839
Device Lot Number0025015987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Initial Date Received by Manufacturer 07/30/2020
Supplement Date Received by Manufacturer09/03/2020
Initial Report FDA Received Date08/26/2020
Supplement Report FDA Received Date09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight70
-
-