Model Number 1977-16-240
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Code Available (3191)
Date of Event 08/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Wrong packaging of the product.No additional information available.Procedure: hip arthroplasty.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : examination of the returned device confirmed the stem is poking through the packaging.The root cause could not be determined.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Event or Problem Description
Additional information received states that there was no adverse consequences that affected the patient and no surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DISTAL TAPERED 16X240A
Common Device Name
RECLAIM IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10454836
Report Number1818910-2020-18798
Device Sequence Number4092544
Product Code KWA
UDI-Device Identifier10603295077183
UDI-Public10603295077183
Combination Product (Y/N)N
PMA/510(K) Number
K102080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number1977-16-240
Device Catalogue Number197716240
Device Lot Number231187
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Initial Date Received by Manufacturer 08/18/2020
Supplement Date Received by Manufacturer10/19/2020
11/10/2020
Initial Report FDA Received Date08/26/2020
Supplement Report FDA Received Date10/26/2020
11/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-