| Model Number |
00884450147627 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that the stent was initially placed (b)(6) 2019.Patient did not attend any follow ups after placement due to alleged incarceration.Patient was seen in (b)(6) 2020 for possible stent removal as the stent had migrated to the stomach.There is no knowledge of when the stent migrated.Dr.Goodman went into retrieve the stent pulling the proximal loop and it had broken off as he was pulling it through the les.The physician stated that he was able to take out about an inch of the stent.At this point the msl was conferenced in to assist.The remainder of the stent had to be piece-mealed through an over-tube until he was finally able to grab the distal loop suture and pull out the rest of the stent into the over-tube.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed; however, the root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|