• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938739
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 07/31/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the clear clips of four (4) 1000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition)bags would not close properly.It was further reported that pliers were used in order to close them.This issue was identified during unspecified process step.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: upon follow-up, the customer clarified that there was no allegation against this product code h938739.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h4 and h6.H4: the lot was manufactured from april 23, 2020 - april 24, 2020.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10457148
MDR Text Key204523698
Report Number1416980-2020-05228
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477237
UDI-Public(01)00085412477237
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2023
Device Catalogue NumberH938739
Device Lot Number60238202
Was Device Available for Evaluation? No
Date Manufacturer Received10/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-