• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674523
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that metal is warped and pin fails to insert and remove properly on a lag screw drill sleeve.Unknown if patient involved.Unknown procedure.Procedure finished with same device.Unknown if there was a backup s&n device.Delay greater than 30 minutes.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device was manufactured in 2014 and shows signs of extensive use.A functional evaluation confirmed that the mating part fails to insert and remove properly, which would cause the device not to function as intended.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRILL SLEEVE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10457564
Report Number1020279-2020-04233
Device Sequence Number17147814
Product Code HSB
UDI-Device Identifier00885556363249
UDI-Public00885556363249
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674523
Device Catalogue Number71674523
Device Lot Number14GCT0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 08/01/2020
Supplement Date Received by Manufacturer12/17/2020
Initial Report FDA Received Date08/26/2020
Supplement Report FDA Received Date12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-