As of today, the medical devices are not available for analysis, therefore the devices themselves could not be investigated.The analysis is thus based on the inspection of the manufacturing documents of the devices as well as on the provided data.The manufacturing process for these devices was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.The analysis of the provided iegm episodes revealed artifacts on the rv channel, which led to the reported oversensing as well as inappropriate shock delivery and icd charging cycle.In spite of the available information, no conclusion can be drawn regarding the root cause of the clinical observation.An analysis of the lead itself would be necessary to determine the root cause.Should additional information or the device itself become available for analysis, the investigation will be updated.
|