MALLINCKRODT MEDICAL SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 86111-
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Codes Extubate (2402); Unintended Extubation (4564)
Date of Event 07/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, the device's cuff pressure was found to be 0.The cuff pressure was checked three times a day before that, and there had been no problem.When the air was injected the pressure was increased the temporarily, but was seen gradually decreased.Re-intubation required.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.Visual inspection noted there was a leak in the inflation line with a clean cut in the material.An inflation/deflation test was performed.25cc of air was applied to the cuff using a syringe and it was observed the cuff deflated after few seconds.It was reported that there was a leak in the cuff without an accompanying report on whether the customer attempted inflation.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: as these devices may have been subjected to handling, storage conditions or preparation which compromised functional integrity; each tube' cuff, pilot balloon and valve should be tested by inflation prior to use.If dysfunction is detected in any part of the inflation system, the tube should not be used.Initiating treatment using a tube already shown to have a dysfunction in the inflation system could unnecessarily subject the patient to the untoward effects of extubation, re-intubation, or loss of respiratory support.Furthermore, the integrity of the inflation system should be monitored both initially and periodically during the intubation period.Uncorrected failure of the inflation system could result in death.Various bony anatomical structures (e.G., teeth, turbinates) within the intubation routes or any intubation tools with sharp surfaces present a threat to maintaining cuff integrity.Care must be taken to avoid damaging the thin-walled cuffs during insertion which would create the need to subject the patient to the trauma of extubation and re-intubation.If the cuff is damaged, the tube should not be used.Do not overinflate cuff.Ordinarily, the cuff pressure should not exceed 25 cm h20.Carroll and grenvik recommend maintaining a seal pressure at or below 25 cm h20 (carroll, r.G., and grenvik, a.: proper use of large diameter, large residual cuffs.Critical care medicine vol.1, no.3:153-154, 1973).Overinflation can result in tracheal damage, rupture of the cuff with subsequent deflation, or in cuff distortion which may lead to airway blockage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key10457651
Report Number8020889-2020-00053
Device Sequence Number3738296
Product Code BTR
UDI-Device Identifier00732094089417
UDI-Public00732094089417
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871204
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86111-
Device Catalogue Number86111-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2020
Initial Date Received by Manufacturer 08/07/2020
Supplement Date Received by Manufacturer11/10/2020
Initial Report FDA Received Date08/26/2020
Supplement Report FDA Received Date11/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-