• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1780KL
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
After battery installation, an "insert a new aa battery" message immediately appeared.After further review, did not note any evidence of previous moisture presence or corrosion on any of the electronic boards, assemblies, or the motor inside the unit.After taking the boards out of the case and installing them into another known good case, the error message appeared again.After swapping a known good pcba2 onto pcba1 and putting them back into the known good case, the error message finally disappeared.The problem seems to be isolated to pcba2.Unable to perform the displacement, sleep current measurement, active current measurement, and self tests due to the recurring error message.(b)(4).
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had failed battery test alert.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
670G INSULIN PUMP MMT-1780KL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
MDR Report Key10458114
Report Number2032227-2020-168776
Device Sequence Number3218444
Product Code OZP
UDI-Device Identifier00763000365882
UDI-Public(01)00763000365882(17)230701
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/01/2023
Device Model NumberMMT-1780KL
Device Catalogue NumberMMT-1780KL
Device Lot NumberHG4FD6H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2020
Initial Report FDA Received Date08/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
Patient Weight125
-
-