| Model Number |
N/A |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was unable to be confirmed with photos received.Photos show size difference of explanted product to correct revision product.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing deficiency.A corrective action has been initiated for further investigation and root cause determination.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial hip arthroplasty.An x ray was taken in post-op, and an abnormality was seen.The doctor decided to re-open the patient and explant the current dual mobility bearing within.After explanation, it was confirmed with calipers that the implanted product was 42mm however the description on the box stated it was a 38mm implant.A new implant was opened and confirmed at 38mm and was implanted into the patient.No further event information available at the time of this report.
|
| |
|