| Model Number |
799-EGB2-01-01 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "runtime calibration failure - 1051" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.The device was put through extensive testing including bench handling, internal inspection, tce/rce functionality testing, and stress testing, with test accessories without duplicating the report.The smart pneumatic module was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|