• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/05/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation could not identify a product problem.The cause of the event could not be determined.This event occurred in (b)(6).
 
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys tsh assay on two cobas 8000 e 801 module analyzers.The sample also had a discrepant result for the elecsys tsh assay ver.2 when tested on the second e 801 analyzer.This medwatch will apply to the elecsys tsh assay ver.2.Patient identifier (b)(6) for information related to the elecsys tsh assay.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2020.The sample was repeated on an abbott architect.The value from the architect analyzer was believed to be correct.The sample was also provided for investigation, where it was tested on a second e 801 analyzer.The e 801 analyzer used at the customer site is serial number (b)(4).The e 801 analyzer used for investigation is serial number (b)(4).Elecsys tsh assay ver.2 reagent lot number 440704, with an expiration date of march 2021 was used on this analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10459931
MDR Text Key206161292
Report Number1823260-2020-02116
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model NumberTSH
Device Catalogue Number08443432190
Device Lot Number440704
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/06/2020
Initial Date FDA Received08/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-