• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 15MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 15MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020 -03377.
 
Event or Problem Description
It was reported that during warehouse inspection, debris was found in the sterile packaging there was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was returned and evaluated against the reported event.Evaluation of the returned product/photographs provided confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.The reported event is confirmed.Dhr was reviewed and no discrepancies were found.The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is that operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 15MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10460679
Report Number0001825034-2020-03378
Device Sequence Number2862302
Product Code PBI
UDI-Device Identifier00880304527447
UDI-Public(01)00880304527447
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000996
Device Lot Number6793729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/06/2020
Supplement Date Received by Manufacturer11/04/2020
Initial Report FDA Received Date08/27/2020
Supplement Report FDA Received Date11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-