|
(b)(4).A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.The unit did not have a battery installed when received.Unit passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and dat test at 0.08730 inches.The stop (idle) current and run current measurement tests are within specification.Unit also passed self test, off no power alarm test and a21 error test.Unit uploaded properly using carelink.Unit had cracked case at display window corner, cracked battery tube threads and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.Data analysis: 07/14/20 00:00:00 daily insulin total = 28.375u, 07/15/20 00:00:00 daily insulin total = 0.000u, 07/16/20 00:00:00 daily insulin total = 0.000u, 07/17/20 00:00:00 daily insulin total = 0.000u, 07/18/20 00:00:00 daily insulin total = 0.000u, 07/19/20 00:00:00 daily insulin total = 0.000u, 07/20/20 00:00:00 daily insulin total = 0.000u.
|