Model Number 72290125
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery, the disposable kit for 1.8mm q-fix mini suture anchor drill and drill guide were inserted into shoulder joint for bicep tenodesis and as drill was advanced, drill became lodged in guide and the drill snapped above the guide, close to the plastic stopper; the procedure was successfully completed without delay using the same device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the device used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A review of risk management files found that the reported failure was documented appropriately.A complaint history review concluded this was an isolated event.A review of the instructions for use found: 1) prior to use, examine the instrument(s) and check for proper functioning.2) the instrumentation is designed for use by surgeons experienced in the appropriate specialized procedures.It is the responsibility of the surgeon to become familiar with the proper techniques.3) additional precautions include those applicable to any surgical procedure.In general, careful attention must be paid to asepsis and avoidance of anatomical hazards.Without the reported product a visual and functional evaluation cannot be performed and customer¿s complaint cannot be confirmed.Factors that could have contributed to the reported event include: (1) excessive force.No containment or corrective actions are recommended at this time.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QFIX 1.8 MINI SUTURE ANCHOR DISP KIT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10462507
Report Number3006524618-2020-00688
Device Sequence Number2199288
Product Code NBH
UDI-Device Identifier00885556662632
UDI-Public00885556662632
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2021
Device Model Number72290125
Device Catalogue Number72290125
Device Lot Number2022649
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/05/2020
Supplement Date Received by Manufacturer09/24/2020
Initial Report FDA Received Date08/27/2020
Supplement Report FDA Received Date09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-