MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM JET7 REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM JET7 REPERFUSION CATHETER; NRY Back to Search Results
Model Number 5MAXJET7-B
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the right m1 segment of the middle cerebral artery (mca) using a neuron max 6f 088 long sheath (neuron max), penumbra system jet 7 reperfusion catheter (jet7), and penumbra system 3max reperfusion catheter (3maxc).During the procedure, the physician advanced the jet7 over the 3maxc and through the neuron max to the target vessel.The physician then completed three passes using the jet7.After the third pass, the jet7 was removed and flushed on the back table.After flushing, the physician noticed that the distal tip of the jet7 expanded and; therefore, the jet7 was not used for the remainder of the procedure.The procedure was completed using a non-penumbra stent retriever with the same neuron max and 3maxc.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #01 mfr report:3005168196-2020-01295 1.Section b.Box 3.Date of event h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JET7 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10462577
Report Number3005168196-2020-01295
Device Sequence Number2147252
Product Code NRY
UDI-Device Identifier00815948020955
UDI-Public00815948020955
Combination Product (Y/N)Y
Initial Reporter CountryVM
PMA/510(K) Number
K190010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/22/2022
Device Model Number5MAXJET7-B
Device Catalogue Number5MAXJET7
Device Lot NumberF93838
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date08/27/2020
Supplement Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Date Report Sent to FDA01/01/2005
-
-