ST. JUDE MEDICAL ADVISOR VL, SENSOR ENABLED CIRCULAR MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
D-AVSE-DF20-V1525 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
08/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, vascular perforation is an inherent risk of any electrode placement.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 3008452825-2020-00444.At the end of a pulmonary vein isolation ablation procedure, a pericardial effusion occurred.The ablation procedure was completed with no reported issues, but at the end of the procedure, after all the catheters were removed, the patient became hypotensive.An echocardiogram revealed a pericardial effusion, for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
|
| |
|