• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 10000425
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was returned and examined and the root cause of the failure could not be established.Review of the device history record notes no material non-conformance¿s, no manufacturing errors, nor discrepancies with respect to material type, treatments, dimensions that may have caused or contributed to this mode of failure.Even though root cause could not be confirmed, the fracture of the tap driver was consistent with off-angled and/or excessive force possibly from sclerotic bone.Labeling review: ".Warnings, cautions and precautions: the implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.".
 
Event or Problem Description
On (b)(6) 2020, patient underwent a two part procedure.First an extreme lateral interbody fusion at l2/3 levels and second, same day a percutaneous posterior fixation.It was reported that a tap driver fractured when a physician was inserting it at the l4 level.The physician tried to remove the piece, however he could not retrieve it.The physician changed the procedure from the percutaneous pedicle screw to an open procedure where the piece was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk blvd.
san diego CA 92121
MDR Report Key10462657
Report Number2031966-2020-00170
Device Sequence Number2344859
Product Code NKB
UDI-Device Identifier00887517511225
UDI-Public887517511225
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160989
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10000425
Device Lot NumberNG79401
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2020
Initial Report FDA Received Date08/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-