Model Number 8PERC
Device Problem Disconnection (1171)
Patient Problem Death (1802)
Event Date 07/22/2020
Event Type  Death  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, an rt was requested to check on the patient noting that the device had an air leak.The patient was not getting good tidal volumes on the ventilator.Spo2 dropped first to the 70s, and then down into the 40s.When the patient was moved from prone to supine position, it was found out that the tracheostomy tube had broken off of the flange, and was nearly 180 degrees inverted in the wrong direction.A replacement of the tube has been done to the patient.But eventually, the patient was reportedly died.The death was not related to the device.
 
Manufacturer Narrative
The incident device has been received and is under evaluation.When the device evaluation is complete a follow-up report will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the flange separated.It was reported that the flange separated from the tube.The reported issue was confirmed.The most likely cause was traced to component issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key10462713
MDR Text Key204637120
Report Number2936999-2020-00634
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
PMA/PMN Number
K884730
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/23/2024
Device Model Number8PERC
Device Catalogue Number8PERC
Device Lot Number19D0635JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Date Manufacturer Received10/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient Weight97
-
-