• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HI FLEX ISRT SZ 5-6 13; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HI FLEX ISRT SZ 5-6 13; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER Back to Search Results
Model Number 71421517
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348)
Date of Event 04/07/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that genesis ii ps hi flex insert size 5-6 13 post broke off inside of the patient 7 years after implantation.Revision surgery was performed but no information about this procedure was initially provided.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per the complaint details, a revision was performed due to a gii ¿poly post broke off¿ approximately 7 years post implantation.S+n has not received the device and/or adequate documentation to fully evaluate the root cause of the reported event; however, no injury was reported.The patient impact beyond the reported post breakage and revision could not be determined.Should additional information/ documentation become available, the clinical/medical task may be re-opened for further evaluation.No further medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT SZ 5-6 13
Common Device Name
PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10463942
Report Number1020279-2020-04271
Device Sequence Number2553223
Product Code JWH
UDI-Device Identifier03596010492951
UDI-Public03596010492951
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71421517
Device Catalogue Number71421517
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/29/2020
Supplement Date Received by Manufacturer06/27/2021
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-