MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX22526UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure, an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a moderately calcified lesion located in the left anterior descending (lad) artery.The device was inspected with no issues noted.Negative prep was not performed.The lesion was not pre-dilated.Resistance was encountered when advancing the device.It was reported that the stent failed to cross the calcified lesion and stent dislodgement occurred during delivery to / at a lesion.The dislodged stent was not removed and left in the patients body.It was stated that it is not known if the stent passed a previously deployed stent, however, another stent was observed in the same vessel.The procedure was completed by pre-dilation and deployments of two non-medtronic stents.The patient was reported to be alive with no injury.
 
Additional Manufacturer Narrative
Additional information: proximal left anterior descending (lad) artery.The lesion was pre-dilated.No attempts were made to remove the dislodged onyx stent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: three procedural images were received.In the first image a guide wire can be seen through a previously deployed stent.Another stent can be seen in the proximal vessel looks like it is in the left main (lm).In the second image diffuse disease due to in-stent restenosis can be seen with a de novo disease/lesion also present.The third image shows the inflation of a balloon inside the previously deployed stent.It appears that the in-stent restenosis and moderate calcification is impacting concentric balloon expansion.This appears to be pre-dilation of the previously deployed stent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10464189
Report Number9612164-2020-03201
Device Sequence Number2642679
Product Code NIQ
UDI-Device Identifier00643169556744
UDI-Public00643169556744
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2021
Device Model NumberRONYX22526UX
Device Catalogue NumberRONYX22526UX
Device Lot Number0010042863
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/26/2020
Supplement Date Received by Manufacturer08/28/2020
11/25/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date09/22/2020
11/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight78
-
-