• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETLW1616C93E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Occlusion (1984)
Date of Event 08/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant iis bifurcate stent graft system was implanted in the endovascular treatment of a 99mm abdominal aortic aneurysm.It was reported that on the date of a ct, three years and four months later, a ia and ib endoleak were observed with aaa sac expansion.Additionally it appeared that the right iliac artery was also occluded.Intervention was performed two days later.An endurant aortic cuff was implanted distally to the origin of the sma to extend the proximal landing zone.An endurant limb was implanted in the right external iliac artery in a good position.The grafts were then ballooned with a reliant balloon.A final angiogram performed demonstrated resolution of the ia and ib endoleaks.As per the physician the cause of the event was anatomy and disease progression.No additional clinical sequelae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Evaluation summary: the reported type ia and ib endoleaks were confirmed on the films provided.The occluded right iliac artery could not be confirmed.Lack of pre-implant ct¿s did not allow for a thorough assessment of the pre-implant anatomy or of the stent graft in vivo positioning at the index procedure.It is likely that disease progression with aortic neck expansion over the 3-4 year implant period was a contributory factor to the observed events.It is also possible that the patient¿s calcified aortic anatomy may have been a factor in lack of seal in the aortic neck and iliac artery.Although iliac occlusion could not be ruled out there appeared to be contrast flow visible on the ct¿s provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10464431
Report Number9612164-2020-03211
Device Sequence Number2642684
Product Code MIH
UDI-Device Identifier00643169467569
UDI-Public00643169467569
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/21/2017
Device Model NumberETLW1616C93E
Device Catalogue NumberETLW1616C93E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2020
Initial Date Received by Manufacturer 08/12/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date10/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-