Model Number 000724
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 06/2023).
 
Event or Problem Description
It was reported that prior to feeding tube placement, foreign material was allegedly found in the port kit.There was no patient contact.
 
Event or Problem Description
It was reported that prior to feeding tube placement, foreign material was allegedly found in the port kit.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review:the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: one tri-funnel replacement gastrostomy tube was returned for evaluation.Gross visual evaluation was performed.The investigation is inconclusive for presence of foreign material issue as hair follicle was noted within the opened packaging, but external to the device and the exact circumstances at the time of event cannot be replicated.A definitive root cause could not be determined based upon available information.Labeling review:a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2023),g4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE, 24F
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10464921
Report Number3006260740-2020-03082
Device Sequence Number3177525
Product Code KNT
UDI-Device Identifier00801741037177
UDI-Public(01)00801741037177
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000724
Device Catalogue Number000724
Device Lot NumberNGDW1641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Initial Date Received by Manufacturer 08/05/2020
Supplement Date Received by Manufacturer11/30/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date12/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No