Model Number 500AHCT
Medical Device Problem Codes Smoking (1585); Sparking (2595)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service representative (fsr) verified the arc on the power cord connection.He did not find any damages due to the sparking and smoking of the power cord.The end user's biomedical engineer changed out the power cord on the bpm.The hlm operated to the manufacturer's specification.
 
Event or Problem Description
After use of the device for a cardiopulmonary bypass (cpb) procedure, the user reported that there was smoke and spark coming from the power cord of the blood parameter monitor (bpm) plugged into the heart lung machine (hlm).There were no reported adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Common Device Name
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10465267
Report Number1828100-2020-00329
Device Sequence Number3092611
Product Code DRY
UDI-Device Identifier00886799001622
UDI-Public(01)00886799001622(11)120828
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer10/06/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date10/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-