| Model Number |
500AHCT |
| Medical Device Problem Codes |
Smoking (1585); Sparking (2595)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service representative (fsr) verified the arc on the power cord connection.He did not find any damages due to the sparking and smoking of the power cord.The end user's biomedical engineer changed out the power cord on the bpm.The hlm operated to the manufacturer's specification.
|
| |
|
Event or Problem Description
|
|
After use of the device for a cardiopulmonary bypass (cpb) procedure, the user reported that there was smoke and spark coming from the power cord of the blood parameter monitor (bpm) plugged into the heart lung machine (hlm).There were no reported adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and or conclusion, a supplemental report will be filed accordingly.
|
| |
|