DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Failure to Pump (1502); Failure to Sense (1559)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to evaluate the iabp unit.The stm connected the ecg leads to the mini patient simulator and the unit display the ecg signal.The stm also connected the simulator and the iabp displayed the ecg/ap.The stm began to run the iabp on a test catheter using the ecg signal as the trigger.The logs were checked and the stm verified the trigger was set to ecg and there were two instances of "no trigger" on the event date.The stm ran the test catheter on ecg trigger mode for one hour and was unable to reproduce the issue.The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
|
|
Event Description
|
It was reported that when the cardiosave intra-aortic balloon pump (iabp) was connected to a patient, it would not pump and would not pick up the ekg.No patient harm, serious injury or adverse event was reported.
|
|