| Model Number |
6531 |
| Medical Device Problem Codes |
Break (1069); Difficult to Advance (2920); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported a shaft break occurred: a percutaneous coronary intervention was being performed on a 75% stenosed, moderately calcified and moderately to severe tortuous lesion.A 12x3.00mm nc quantum apex balloon was difficult to track over the guidewire and the shaft of the balloon kinked.During pullback of the balloon the shaft fractured partly.The balloon was completely recovered and removed from the patient.The procedure was successfully completed with a different device without issue or patient injury.There were no adverse patient effects and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of an nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 79.1cm distal of the strain relief.There was contrast and blood in the inflation lumen.The balloon was loosely folded.The device was functionally tested with a.014 inch guidewire and no issues were detected.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
|
| |
|
Event or Problem Description
|
|
It was reported a shaft break occurred.A percutaneous coronary intervention was being performed on a 75% stenosed, moderately calcified and moderately to severe tortuous lesion.A 12x3.00mm nc quantum apex balloon was difficult to track over the guidewire and the shaft of the balloon kinked.During pullback of the balloon the shaft fractured partly.The balloon was completely recovered and removed from the patient.The procedure was successfully completed with a different device without issue or patient injury.There were no adverse patient effects and the patient was stable.
|
| |
|
Search Alerts/Recalls
|