MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT1H04
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-01321.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a dural arteriovenous fistula (davf) using penumbra smart coils (smart coils).During the procedure, two smart coils would not advance at all from their initial position within their introducer sheaths.Therefore, the smart coils were removed.It was reported that while attempting to advance one of the smart coils, the pusher assembly of the smart coil inadvertently bent.It is unknown which of the two smart coils pusher assembly was the one that bent.The procedure was completed using other smart coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device in complaint was not expected to be returned; however, on (b)(6)-2020 the device was received.Additional information was added to: 1.Section d.Box 10: returned to manufacturer on (mm/dd/yyyy) results: the pet lock was intact on the proximal end of its pusher assembly.The pusher assembly was kinked approximately 20.0, 69.0, 120.0 and 138.0 cm from its proximal end.The pusher assembly mid-joint was proximal to the introducer sheath friction lock.The embolization coil was intact with its pusher assembly and undamaged.Conclusions: evaluation of the first returned smart coil revealed that the pusher assembly mid-joint was proximal to the introducer sheath friction lock.Further evaluation revealed that the pusher assembly was fractured, and the embolization coil was detached.If the pusher assembly mid-joint is proximal to the introducer sheath friction lock, resistance may be encountered upon advancement, and the smart coil may not advance within its introducer sheath.If the device is forcefully advanced against this resistance, damage such as a kink and subsequent fracture may result.Evaluation of the second returned smart coil revealed that the pusher assembly mid-joint was proximal to the introducer sheath friction lock.If this occurs, resistance may be encountered upon advancement, and the smart coil may not advance within its introducer sheath.If the device is forcefully advanced against this resistance, damage such as a kink in the pusher assembly may result.During the functional test, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.The smart coil was then advanced out of its introducer sheath without an issue.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-01321 h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10467453
Report Number3005168196-2020-01322
Device Sequence Number14203549
Product Code HCG
UDI-Device Identifier00814548015248
UDI-Public00814548015248
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2025
Device Model Number400SMTXSFT1H04
Device Catalogue Number400SMTXSFT1H04
Device Lot NumberF95910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer10/29/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date11/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-