• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71173547
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Date 08/06/2020
Event Type  malfunction  
Event Description
It was reported that the 2.5mm cannulated screw driver shaft became stuck inside of the large screwdriver handle.No case, therefore, no patient was involved.Backup device was available.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the 2.5mm cann hexdriver shaft - 120m is stuck inside the large screwdriver handle and cannot be removed.These devices show signs of significant wear and use.Due to the devices being stuck together, it is difficult to read the lot numbers from the devices.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.These devices are reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE SCREWDRIVER HANDLE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10468527
MDR Text Key204856578
Report Number1020279-2020-04273
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010519580
UDI-Public03596010519580
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173547
Device Catalogue Number71173547
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2020
Date Manufacturer Received08/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-