SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
Model Number 71173547 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/06/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the 2.5mm cannulated screw driver shaft became stuck inside of the large screwdriver handle.No case, therefore, no patient was involved.Backup device was available.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the 2.5mm cann hexdriver shaft - 120m is stuck inside the large screwdriver handle and cannot be removed.These devices show signs of significant wear and use.Due to the devices being stuck together, it is difficult to read the lot numbers from the devices.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.These devices are reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|