• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TX027 TX-JACKSONVILLE-MCKNIGHT CRD FLEX 3000CC CANISTER WITH VALVE; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TX027 TX-JACKSONVILLE-MCKNIGHT CRD FLEX 3000CC CANISTER WITH VALVE; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number 65652-631A
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Codes Cardiac Arrest (1762); Death (1802)
Date of Event 08/09/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
Investigation into this reported incident is on-going while we continue to gather further details related to the event.A follow-up report will be filed once the results have been completed.
 
Event or Problem Description
The suction canister is used without a stand keeping it vertical and can be lying down or shocked at any time during resuscitation.This may result in the safety valve being clamped on the outlet and thereby stopping the suction.Lack of information on the device, on the risk of stopping the vacuum.This type of event was suspected during loss of suction during cpr (cardiopulmonary resuscitation) in emergencies, having resulted in eigas (serious adverse event related to treatment) death of patient.
 
Additional Manufacturer Narrative
Follow up report is being filed, since the results of the investigation are now available.A thorough investigation into the incident reported, has been completed regarding the outer canister used with a disposable liner.Based on the information obtained from the healthcare facility in france.It was determined, that the canister was not placed into a wall bracket or roll stand as stated, in the directions for use.Therefore, the device failed to work as intended.The dfu under setup states, "if not already in a mounting bracket, place reusable hard canister into a bracket on a roll stand or wall mount.Device is to be used in vertical orientation only".Canister shut off and suction stopped since device was used without the holder to maintain vertical position as required.The product worked as designed.Cardinal health¿s manufacturing team will be adding an orientation arrow onto the canister, that will help the end users who do not necessarily see the instructions for use (ifu) on proper setup of the canister, with liner setup to prevent improper setup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRD FLEX 3000CC CANISTER WITH VALVE
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
TX027 TX-JACKSONVILLE-MCKNIGHT
200 mcknight st.
jacksonville TX 75766
MDR Report Key10468687
Report Number1423537-2020-00506
Device Sequence Number3273781
Product Code GCX
UDI-Device Identifier10885380158612
UDI-Public10885380158612
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number65652-631A
Device Catalogue Number65652-631A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2020
Supplement Date Received by Manufacturer08/21/2020
Initial Report FDA Received Date08/28/2020
Supplement Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-