MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC GEL PADS Back to Search Results
Catalog Number 317-09-02
Device Problems Insufficient Flow or Under Infusion (2182); Gas Leak (2946)
Patient Problems No Consequences Or Impact To Patient (2199); Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device was getting low flow (alert 02) alert. Flow rate 1. 8lpm with large pads was in place. Initial inlet pressure was 5. 7psi. Disconnected and reconnected pads using proper technique. Inlet pressure was -4. 4psi. And noted there are a lot of bubbles in the left chest pad. The nurse confirmed that the patient went to computed tomography with pads in place. When disconnected the left chest pad flow rate increased to 2. 1lpm and inlet pressure settled at -7psi. Asked to hold on to the pad for investigation and replaced with a universal pad.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC GEL PADS
Type of DeviceARCTIC GEL PAD
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key10469004
MDR Text Key206141924
Report Number1018233-2020-05539
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number317-09-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-