| Model Number |
TC-XXXXX |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The issue was identified prior to any patient involvement (unpacking).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The involved device has been requested at livanova usa inc for investigation.Investigation is ongoing.If any additional information pertinent to the reported event is obtained, it will be provided in a supplemental report.Device not yet returned.
|
| |
|
Event or Problem Description
|
|
Livanova has received a report that the catheter thoracic 32fr x 20in straight flared end (catalogue number tc-10132, lot 2009000080) contained a thoracic catheter french 28fr instead of french 32fr as per labelling.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint was confirmed through the analysis of provided picture.The dhr review was carried out: the lot was released as conform according to the specifications.No deviation, non-conformity, or material scrap/requests relevant to the reported issue was identified.No further similar complaint in the last 12 months occurred for the claimed code.Based on the internal investigation the most probable root cause for the complained failure was a mix-up problem occurred during kitting process.In particular, the manufacturing personal pulled the wrong product, the 28fr cannula, instead of the 32fr cannula, and allocated those to the shop order.To prevent recurrence of the complained failure, a capa activity has been initiated.Despite the risk level associated to this issue is low, to remove the affected devices from the market, a field action was initiated.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|