| Model Number |
EPS02 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Injury (2348); Bowel Perforation (2668)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unknown.Investigation summary: as the device was not returned, an analysis investigation could not be performed.A conclusion could not be reached as to what may have caused or contributed to the event.Complaint information is trended on a regular basis to determine if further investigation is warranted.The lot/batch/serial number was not provided; therefore, a manufacturing record evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a laparoscopic ob-gyn procedure, the device was activated when the device tip was not touching the target tissue and capacitive coupling had occurred outside the visual field of scope.In this time, the shaft part was touching the intestine and mesentery.The intestine was damaged and ileocecal perforation was found after the procedure.The device was used to complete the case.The patient was discharged from the hospital.No device will be returning.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/10/2020.Additional information was requested and the following was obtained: were any other energy devices used in this procedure? there is no report about the other energy devices.No further information will be available.Why is it believed that the thermal damaged was caused by the probe plus ii shaft?do they believe the thermal damaged was caused by shaft ? dr.Commented that the shaft part was touching to the intestine and mesentery.No further information will be available.What were the power setting on the generator during the procedure? no further information will be available.Approximately how long was device activated during the surgical procedure? no further information will be available.What heat management techniques were used throughout the procedure? no further information will be available.How was the patient's condition managed? no further information will be available.What is the patient¿s current status? she is already been discharged from the hospital.No further information will be provided.".
|
| |
|
Search Alerts/Recalls
|