| Model Number |
364508 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The orsiro drug-eluting stent system was chosen for the treatment of a calcified lesion.After pre-dilatation and implanting of two other orsiro stents, the complaint device could not pass the lesion.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned device was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon is well folded and shows no signs of inflation.Judging from the x-ray markers the stent is not displaced.The stent does not show any damage or irregularities.The crimped diameter of the stent still complies with the specification.Review of the production documentation of the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The root cause for the complaint event is most likely related to the patients anatomy (i.E.Previously implanted stent).
|
| |
|
Search Alerts/Recalls
|