Model Number 364508
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/04/2020
Type of Reportable Event Malfunction
Event or Problem Description
The orsiro drug-eluting stent system was chosen for the treatment of a calcified lesion.After pre-dilatation and implanting of two other orsiro stents, the complaint device could not pass the lesion.
 
Additional Manufacturer Narrative
Combination product: yes.The returned device was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon is well folded and shows no signs of inflation.Judging from the x-ray markers the stent is not displaced.The stent does not show any damage or irregularities.The crimped diameter of the stent still complies with the specification.Review of the production documentation of the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The root cause for the complaint event is most likely related to the patients anatomy (i.E.Previously implanted stent).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.0/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10473351
Report Number1028232-2020-03679
Device Sequence Number3273517
Product Code NIQ
UDI-Device Identifier07640130439177
UDI-Public7640130439177
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2021
Device Model Number364508
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10194619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Initial Date Received by Manufacturer 08/28/2020
Supplement Date Received by Manufacturer10/14/2020
Initial Report FDA Received Date08/31/2020
Supplement Report FDA Received Date10/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-