BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10604
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/09/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 100% stenosed target lesion was located in the left anterior descending artery.A 2.25 x 24mm synergy ii drug eluting stent was advanced for treatment.However, during insertion, the shaft was fractured.The device was removed with the wire and the procedure was completed with a new stent.There were no patient complications reported.
 
Event or Problem Description
It was reported that shaft break occurred.The 100% stenosed target lesion was located in the left anterior descending artery.A 2.25 x 24mm synergy ii drug eluting stent was advanced for treatment.However, during insertion, the shaft was fractured.The device was removed with the wire and the procedure was completed with a new stent.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy ii us mr 2.25 x 24 mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured using snap gauge and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube shaft found a break situated at 16.2 cm distal to the distal end of the strain relief as well as multiple hypotube kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10474021
Report Number2134265-2020-11173
Device Sequence Number19567834
Product Code NIQ
UDI-Device Identifier08714729840206
UDI-Public08714729840206
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/07/2021
Device Model Number10604
Device Catalogue Number10604
Device Lot Number0024097621
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date08/31/2020
Supplement Report FDA Received Date10/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-